<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\NVSStyleSheet.xsl"?>
<Parameter>
<ParameterFamily>head</ParameterFamily>
<LongName>
humidity
</LongName>
<Name>
humidity
</Name>
<AppliesTo>
<UnnumberedList>
<ListItem><TextApplies>Thermal</TextApplies></ListItem>
</UnnumberedList>
</AppliesTo>
<Description>
<DescriptionText><text>
The thermal head firmware monitors the humidity using a sensor mounted within
the thermal head.  If the humidity exceeds the configured threshold, then the
thermal head is shutdown.  Any plot in progress is aborted.
</text></DescriptionText>
<DescriptionText><text>
If this parameter is set to zero, then the humidity sensor is ignored by the
thermal head firmware. This value should never be set to zero on any head in
the field since water condensation on the laser will cause laser failure.
</text></DescriptionText>
</Description>
<Type>
real
</Type>
<Default>
<UnnumberedList>
<li>30.000</li>
</UnnumberedList>
</Default>
<Range>
<UnnumberedList>
<li>0.000 to 60.000 [%]</li>
</UnnumberedList>
</Range>
<ArraySize>
</ArraySize>
<Notes>
<DescriptionText><text>
Applies to Thermal Head firmware version 1.30 and above
</text></DescriptionText>
</Notes>
<Partition>
cal
</Partition>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<li><Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link></li>
<li><Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link></li>
<li><Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link></li>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:37:00 2008 PST
</Updated>
<Links>
</Links>
</Parameter>